注射器微量注射泵可用于動物實驗,對動物進行微量注射,可控制流速, 可通過程序控制注射流量。
主要性能特征:
??
??? 1. 單機操作也可電腦控制 (RS-232插頭)
????2. 注射也可回抽 (雙向)
??? 3. 用1~60cc注射器
??? 4. 可選擇注射流量及流速控制程序,有41種選擇
????5. 多個泵連接組合成一個系統 (雙筒), 最大可達100泵;
??? 6. 單筒NE-1000、6筒NE-1600、8筒NE-1800整機選擇.
??? 7. 注射速率pumping rate:
?????? NE-1000: 0.73?? ml/小時~2100ml/小時
?????? NE-1600: 0.568 ml/小時~1337ml/小時
?????? NE-1800: 0.568 ml/小時~380? ml/小時
?
?
規格表Descriptions
- Model :?? ??????????NE-1000????????? NE-1600?????????? NE-1800
- Syringe sizes:?? 1 ~ 60 cc 1 ~ 60 cc????????? 1 ~ 10 cc
- Number of syringes:?????? 1~ 6 8
- Electrical
- Motor type: Step motor
- Motor steps per revolution: 200
- Motor to drive screw ratio: 1/4
- Drive screw pitch: 24 revolutions
- Advance per motor step: 1.323 um
- DC connector: 2.5 mm, center positive
- Voltage at DC connector: 12V DC at full load
- Amperage: 1000 mA at full load
- Power supply output rating: 12V DC @ 1000 mA
- Dimensions:
- NE-1000: 22.86 cm x 14.605 cm? x 11.43 cm
- NE-1600,1800: 26.35 cm x 38.42 cm? x 13.34 cm
- Weight: NE-1000 (1.63 kg)? NE-1600,1800? (4.6 kg)
- Operational
- Maximum speed:? 3.2226 cm/min?????????????
- Minimum speed:?? 0.0262 cm/hr
- pumping rate:
?????? NE-1000: 0.73?? ml/hr~2100ml/hr
?????? NE-1600: 0.568 ml/hr~1337ml/hr
?????? NE-1800: 0.568 ml/hr~380? ml/hr
- Maximum force: 160 lbs at minimum speed
?????????????????????????????????????? 30 lbs. at maximum speed
- Number of Program Phases: 41
- RS-232 pump network:? 100 pumps maximum
- Syringe inside diameter range: 0.100 ~ 50.00 mm
|
|
? |
- Infuses and withdraws
- Selectable pump rate units: μl/hr, μl/min, ml/hr, ml/min.
- Change pumping rate and direction while pumping.
- Dispense a specified volume.
- Syringe purge mode will infuse or refill syringe at top speed to purge air from the line or to fill the syringe.
- Audible buzzer can be programmed to alert when an alarm condition occurs or the pumping program completes.
- Power failure mode restarts a pumping program interrupted by a power failure.
- Stall detection stops pump when the motor's operation is impeded (not on NE-500 model)
- Non-volatile memory restores all setup parameters and the pumping program, on power up.
- Elastomer raised keypad with optional beeping whenever the keypad is pressed.
- CE certified
|
|
|
? |
- Pre-program a set of production dispensing protocols. The operator then selects the required dispense protocol
- Keypad Lockout: Prevent unauthorized or accidental changes to the settings by requiring a "Lockout Disable" key to be inserted to change settings.
|
|
|
? |
- 41 programmable phases
- Enter programs from the keypad or from a computer.
- Download or upload pumping programs from or to a computer. Store dispensing programs in a file on your computer, then download them to one or more pumps. Specify in the file to send a different pumping program to each pump in the pump network.
- Pre-program dispense volumes. Each volume can be different.
- Automatically change pumping rates or pumping direction.
- Ramp up or down the pumping rate.
- Set timed delays between dispenses.
- Automatically pause the program and wait for the user to continue the dispense.
- Synchronize dispenses with other equipment or pumps to change pumping rates in reaction to a sensor or signals from other pumps.
- Send logic signals to other equipment. Change pumping rates in reaction to a sensor.
- Program the audible alarm to beep at any time to alert the operator.
|
|
? |
|
|
? |
|
|
- Network up to 100 pumps from a single communications port on a computer. Baud rates up to 19200.
- Free terminal emulator program allows communications with syringe pump.
- Directly control one or more pumps from a computer, including setting the pumping rate, direction and volume. Set the operational and setup configurations. Also the TTL interface can be queried and the programmable output set.
- Using a computer, each pump in the pump network can be individually controlled and monitored. Or, the computer can download a new pumping program to each pump which would automatically reconfigure the dispensing system for a new job.
- Downloading control programs to the pump would unload computer overhead to the pump. Computer resources could then be used elsewhere while the pumps operate independently.
- 2 modes of communications: Basic and Safe
- Basic: Basic master-slave interface. Easy to implement. Can communicate from any terminal emulator.
- Safe: A reliable interface for a real-world environment.
- - High quality interface which detects communication errors via a 16 bit CCITT CRC.
- - The pump will automatically transmit an alarm packet when a pump alarm occurs. For instance, if the pump stalls or resets, a message is immediately sent to the computer.
- - A configurable communications time-out will stop the pump and alarm if the communications link is broken.
- - Packet inter-character time-out to recover packet synchronization after communication glitches.
- Software Developer friendly communications protocol.
- Consistent and well defined communications packets.
- Constant start and end characters.
- Packets structured to be easily parsed
|
|
? |
|
|
? |
|
|
- Control the pump from a foot switch or timer.
- Output signals can be used to switch a valve from an output to an input reservoir.
- Keypad Lockout: Prevent unauthorized or accidental changes to the settings by requiring a "Lockout Disable" key to be inserted to change settings.
- Pumping direction control
- Programmable output signal
- Programmable responses to input signals
- Configurable start and stop trigger:
- Start or stop the pump on each contact closure to ground. For example, a foot switch.
- Start the pump on closure to ground and stop the pump on opening the connection. For example a timer relay that starts the pump whenever closed and stops the pump whenever open.
- Only start the pump on each contact closure to ground. For example, a laboratory animal trained to press a lever can receive a drug dosage from the pump and the pump will not stop or deliver another dosage until it is permitted by the pumping program entered by the user
- Configurable trigger input:
- Edge trigger: Starts or stops the pump, such as from a foot switch
- Level trigger: Start or stops the pump, such as from a timer or relay control
- Start only trigger: Edge trigger will only start the pump, such as from a lab animal trained to press a lever to receive a dosage.
? |
|
|
|
Special Application: 雙筒連續灌注Continuous Infusion/Dual Syringe Pump
?
? 注射器加熱裝置
?
??
??
SYRINGE-HEATER-KIT |
注射器加熱裝置 |
Description:
- Heating device for syringes that require temperature-controlled dispensing
- Heats syringe to over 100 degrees Celsius
- Digitally set the heater setpoint
- Temperature sensor in heating pad
- Setpoint retained in memory
- Controller will heat the syringe and hold the set temperature using an on/off or PID control algorithm
- Multiple syringe heater pads can be attached to the control unit
- Heated syringe can be mounted on a syringe pump
- Syringe Heater Kit includes Control-Unit power cable and Primary-Syringe-Heater-Pad
|
? |